Dovient

How-to

Medical & biomedical equipment maintenance program

How to set up a medical equipment maintenance program for an ambulatory surgery center: the inventory, the schedule, who does the work and the records surveyors check.

What the rules expect

Medicare requires an ASC's emergency equipment to be immediately available, suited to its patients and maintained by appropriate personnel (42 CFR 416.44(d)), and the building to meet NFPA 99, the Health Care Facilities Code (416.44(c)). Chapter 10 of NFPA 99 covers electrical safety of patient care equipment, including testing before first use and after repairs, with ongoing testing set by facility policy. Surveyors expect each device to be maintained on the manufacturer's recommended schedule, unless your accreditor allows a documented alternate program.

Biomedical equipment maintenance is usually split: clinical staff do daily checks (defibrillator self-test, crash cart, suction), and a biomed vendor or in-house biomed tech does the preventive maintenance, calibration and electrical safety tests.

How to set up the program

  1. Build the inventory: every device with its make, model, serial number, location and the manufacturer's PM interval. An ASC equipment list usually covers anesthesia machines, patient monitors, defibrillators, electrosurgical units, sterilizers, suction, infusion pumps, scopes and OR tables.
  2. Do an incoming inspection on every new or loaner device before it touches a patient, and add it to the inventory.
  3. Set each device's PM interval from the manual, and its electrical safety test interval from your policy under NFPA 99.
  4. Assign each task: staff daily checks, vendor PMs, sterilizer monitoring.
  5. Tag each device with its last and next PM date, and take any device that fails out of service with a tag.
  6. Check FDA recalls and manufacturer notices against your inventory, and record what you did.
  7. Review the program yearly: overdue PMs, repeat failures, devices to replace.

A typical maintenance schedule for an ASC

  • Daily, before cases: defibrillator check, crash cart seal and log, suction, anesthesia machine checkout.
  • Each sterilizer cycle and as the manufacturer and your policy set: chemical and biological monitoring.
  • On the manufacturer's interval, often every 6 or 12 months: vendor PM and calibration for each device.
  • On your NFPA 99 policy: electrical safety tests, and always after a repair.
  • Yearly: inventory audit and program review.

Records to keep

For each device: the PM and electrical safety reports with dates and who did them, repairs and parts, daily check logs, incoming inspection, recall actions and the date it was retired. Surveyors pick devices in the OR and ask to see their PM history, so the record for each serial number needs to be complete.

Run it in Dovient Lite

Add each device as an asset with its serial number and location, or import the inventory from a spreadsheet, and print a QR label for each. Set a schedule for each PM at the manufacturer's interval, and daily checklists for the crash cart and defibrillators; a work order opens when each is due, and the vendor's report is attached to it. Each device's record shows every PM, repair, part and cost, and completed work orders export as a PDF for surveyors.

Start with Dovient Lite

Sources

Rules change and vary by state and county. Check the current text with your inspector or authority having jurisdiction. Last reviewed 2026-11-15.

Questions

How do I set up a medical equipment maintenance program?
Build an inventory with each device's PM interval, assign daily checks to staff and PMs to a biomed vendor, tag each device with its due date, and keep a record of every PM, repair and recall action.
How often should medical equipment get preventive maintenance?
At the manufacturer's recommended interval, which is often every 6 or 12 months. Your accreditor may allow a different documented schedule for some devices.
What's a good maintenance schedule for an ambulatory surgery center?
Daily checks of defibrillators, crash carts, suction and anesthesia machines; sterilizer monitoring with each load as your policy sets; vendor PMs on the manufacturer's interval; and a yearly inventory and program review.
What records do surveyors ask for?
The equipment inventory, PM and electrical safety reports for the devices they pick, daily check logs and evidence you act on recalls.