What CDC recommends for sterilization monitoring
CDC's dental infection control guidance uses three kinds of monitoring, and each one belongs in your sterilization log.
- Mechanical: check the gauges, display or printout for time, temperature and pressure on every cycle.
- Chemical: an indicator inside every package, and an outside one if the inside one cannot be seen. If the color did not change, do not use the instruments.
- Biological (spore testing): at least weekly for every sterilizer, even ones used part time, with a matching control from the same lot number.
- Every load with an implantable device gets a spore test, and the implant ideally is not used until the test comes back negative.
Who requires it
CDC's guidance is a recommendation, not a federal rule. Many state dental boards make weekly spore testing a requirement and set how long you keep the records. CDC says to keep sterilization records in line with state and local regulations, so check your state board's rules for the retention period.
What to do after a positive spore test
- Take the sterilizer out of service and review how it was loaded and run.
- Recall and quarantine any implantable items from that load until shown to be sterile.
- Repeat the spore test right away, using the same cycle that failed.
- If the repeat test is negative and procedures were correct, put the sterilizer back in service.
- If the repeat test is positive, keep it out of service until it is inspected or repaired and passes spore tests in three consecutive fully loaded cycles.
- Where possible, recall and reprocess items from suspect loads back to the last negative test.
Autoclave maintenance to log
Many failed tests come from the sterilizer, not the instruments. Follow the manufacturer's manual, which usually calls for daily or weekly chamber cleaning, checking the door gasket, draining and refilling the water reservoir with distilled water, and cleaning the chamber filter. Log each of these and any service visit next to the spore tests, so a pattern of failures can be traced to its cause.
Autoclave sterilization and spore test log
Every cycle, with a spore test at least weekly and for every implant load
| Date | Sterilizer | Load number | Time, temp and pressure OK | Chemical indicator (pass or fail) | Spore test result | Control result | Initials |
|---|---|---|---|---|---|---|---|
The printed log has 10 rows per page.
Run it in Dovient Lite
Add each autoclave as an asset and set a weekly spore test schedule with a short checklist (test vial run, matching control run, result negative). A work order opens each week; the assistant marks the result pass or fail on the phone and attaches a photo of the lab report. Set the manufacturer's cleaning and gasket checks as their own schedules, and the completed checks export as a PDF for a state board inspection.
Start with Dovient LiteSources
- CDC, Dental infection control: sterilization and disinfection
- CDC, Dental infection control FAQs: sterilization monitoring
Rules change and vary by state and county. Check the current text with your inspector or authority having jurisdiction. Last reviewed 2026-11-15.
Questions
- How often should a dental office do spore testing?
- CDC recommends a spore test at least once a week for each sterilizer and for every load that contains an implantable device. Many state dental boards require weekly testing.
- How long do I keep spore test records?
- CDC says to follow your state and local rules, and the period differs by state. Check your state dental board's regulations.
- What happens if a spore test fails?
- Take the sterilizer out of service, review how it was used, and repeat the test right away. If it fails again, the sterilizer stays out until it is repaired and passes three consecutive tests.
- What goes in an autoclave log sheet?
- The date, the sterilizer, the load, whether time, temperature and pressure were correct, the chemical indicator result, the weekly spore test and control results, and who ran it.